EMTEK’s P100 Portable Microbial Sieve Sampler is designed to operate in a variety of Critical Environments including LAF Hoods, BSCs, Isolators, Sterility Chambers, Filling Lines, Bulk Production Areas, Surgical Amphitheaters, and Compressed Air/Gas Systems, but are practical for use in almost any environment.
P100 CONTROLLER – General Specifications
- Controlled Sample Rates of 28.3 and 100 LPM
- Dimensions: LxWxH = 7.5" x 5.5" x 6" / Weight: <4.4 lbs (2 kg)
- LCD View Screen w/Touch Screen Interface
- Controller Enclosure: Kydex® with Microban® (for Anti-Microbial Protection)
- P100 Inlet Cover Anodized 6061-T6 Aluminum with color-coded 28.3 or 100 LPM Sample Rates, also available in 316SS.
- Inlet cover is a sieve pattern of 300 holes which are appropriately sized for either 28.3 or 100LPM flow rates
- Media Stage with Adjustable height to accept 15-40ml Fill 90mm Test Plates
- HEPA Filtered Exhaust, 0.2 Micron (w/remote exhaust tubing option)
P100 MICROBIAL AIR SAMPLING PACKAGE - with 6061-T6 Aluminum or Optional 316SS Inlet Cover
Standard Package Includes:
- P100 Air Sampler
- Includes P100 Version 2.00 FW, with Enhanced CFR 21
Part 11 Compliance for P100 - Inlet Cover (Choose 1): 28.3 LPM, or 100 LPM
Note: Specify Desired Inlet Cover Flow Rate(s) when ordering. - IR Remote Control / 5 Channels (Start/Pause/Stop)
- Inlet Cover Lid / Port & Plug Set (USB, RJ45, Inlet/Outlet Plug)
- User Manual & Associated Documents (on 1 8GB USB Drive)
- AC/DC Power Supply/Battery Charger
- NIST Traceable Calibration and Certificate
- Custom Carrying Case (Plastic Case, Custom Component Inserts)-Standard 2-Year Warranty
P100 DATA/COMMUNICATION
- Ethernet Port: Unit Calibration, PC Communication, and Data Output to PDF or Optional Printer
- USB Port: Data Retrieval via USB Drive, or PC
- User Selectable Parameters: Sample Rate, Sample Time or Volume, Delay/Test/Hold, Time Base, Date Format, User ID, etc.
- Sample Run Memory (500 Runs) – Critical Sample Run Parameters Captured
- Sample Run Data Output to Printer, USB, or PC / Sample Label, Locked PDF Report, or .csv File Outputs.
- Data Control through System Admin/User ID Control (Part 11 Compliant)
- Site Description Entry/Maintenance/Capture
- Optional: PC Control Software - Offers Substantial Operative PC Control of a single, or multiple P100's via Ethernet connection to a
Local Area Network (LAN). General Control: Start, Stop, Pause, Resume, Set/View: Date, Time, Sample Volume, Sample Time,
Delay, Test, Hold, Site ID, Bldg/Room Data View/Output: On Screen, .csv File, or Locked PDF Report Output
P100 Optional Accessories
- Stainless Steel Inlet Cover, available for both 28.3 or 100LPM flow rates
- Compressed gas sampling kit, includes filtered High Pressure Diffuser, compressed gas sampling inlet adapter, clamps, and 2.5’ of tubing
- Portable handheld thermal printer kit, including thermal printer, ethernet communication cable, charger, and printer label rolls
- PC Control Software (See data/communication section for details)
- Remote sampling via open-flow assembly to an EMTEK Remote Air Sampler (RAS) attached w/ up to 10’ of tubing
- Remote dual 28.3LPM sampling via barbed inlet cover utilizing tubing with two EMTEK Remote Air Samplers (RAS)
| Sample Time/Volume | Variable (User Defined), Maximums: 120-minutes/3396 Liters* @ 28.3 LPM, 30-min/3000L @ 100 LPM |
| Delay/Hold Times | Variable (User Defined) |
| Sample Flow Rates | 28.3 LPM (1 CFM), or 100 LPM (Requires separate inlet covers for the 28.3 & 100 LPM Sample Rates) |
| Flow Rate Control | Electronic, Closed-Loop, Mass-Flow Control |
| Printer (Optional) | Thermal Label or Paper |
| Control System (CPU) | Microprocessor Controlled (32bit PIC Processor) |
| Memory | 512kb Flash Program , 128kb RAM Data, 1mb Sample Runs, 512kb EPROM Calibration Set Points |
| Unit Equipment ID/Number | User Defined/Selectable |
| Battery Operation | ~6 Hours at 100 LPM & ~10 Hours at 28.3 LPM / ~2.5 Hour Charge Time |
| Site Descriptions | User Created/Deleted/Selectable |
| Program Descriptions | User Created/Deleted/Selectable (Includes: Sample Rate, Volume/Time, Flow/Volume Units, Delay/Test/Hold) |
| Sample ID | Unique System Generated (Unit Serial # + 5 digit string) / Non-Repeating until >999,999 samples |
| Input/Output | USB Client 1.1, Ethernet 10BaseT-/100-BaseT |
| Audible Alarm | Internal (with User Volume Control) |
| Alarms | Flow Alarm +5% (On/Off) |
| AC/DC Power Supply | Input: 100-240 VAC, 50/60 Hz,130VA-168VA 1.4 AMPS / Output: DC 18V 3.6A |
| Operating Range | 5-40º C, 10-80% RH, non-condensing*; Indoor Use; Max Altitude 6560 feet (2000 meters) Note: As temperature increases from 30 to 40º C, humidity range drops from 80 to 50% linearly. |
| Calibration/Verification Required | Flow Rate (28.3/1 CFM and/or 100 LPM) / Sample Timer (Frequency: 6-12 Months) |
| Installation Category/Pollution Degree | Category 1 / 1 & 2 |